56

Standard Operations

5.7.1.1 Set up single wavelength mode

Press Single Wavelength in the Main Menu. Press Options for
Parameter Setup.

Table 7

describes the single wavelength setup

options.

Table 7 Single wavelength setup options

Options

Description

More

For further Options

Store Off/On

With the Store On setting, all measurement data are stored automatically. With the Store Off
setting, no measurement data are stored.

% Trans/Abs

To switch to % transmittance, concentration or absorbance readings

λ

Wavelength

To enter the measurement wavelength. Use the alphanumeric keypad to enter the measurement
wavelength. The entered wavelength must be in the range from
320–1100 nm.

Send Data

To send data to a printer, PC or USB memory stick

Concentration Factor

Multiplication factor for converting absorbance values into concentration values.

Concentration
Resolution

To select the position of the decimal point in the calculated concentration readings.

Reading Mode

Single Reading Mode: A reading is only displayed after a measurement has been carried out
(press Read; standard setting) (see

section 5.7.1.2 on page 58

).

Continuous Reading Mode: After the zero measurement, all readings are displayed
automatically and continuously (see

section 5.7.1.3 on page 58

).

Recall Data

Call up saved measurement data, wavelength scans or time courses, see

section 5.3 on

page 28

.

Instrument Setup

Basic data of the instrument, see

section 5.2 on page 17

.