Standard Operations
5.
Press Options > View Values the color value of the sample is
displayed.
5.6.2 Take a color difference measurement with stored reference values
1.
Press Color Difference Measurement from the Main Menu,
2.
Press Select Reference.
3.
A list of available references is displayed.
4.
Press the required reference.
5.
Press Select reference to confirm.
6.
Insert the sample cuvette/cell.
Important Note: For color difference measurement, it is absolutly
necessary that the type of cuvette/sample cell used for sample
measurment is identical to the type of cuvette/sample cell which
was used for reference measurement.