Calibration Using Converter SignaJ Source

9.2-1

0 % Flow Rate Calibration

■ To check the zero point of flow measurement:

Key operation

Display example

Description

N1;FL0H CAL OX

0

.

0

%

Press [SET] first to start the items selection scquertcc

and select N1: FLOW CAL 0% from among the
configuration items using [ ► ] and [ A ] keys, 'fheii

press [SET] again to go into the calibration mode
and calcuiaie and display zero point using a

simulation signal,

SET

N1:FL0W CAL

Q%

Pressing [SET], the system re turns to the items

selection sequence.

To change the zero and span of the converter:

Key operation

Display example

Description

SET

N l : F L 0 i C A L O X

0

.

1

%

Press [SET] first to start the items selection sequence

and select N1: FLOW CAI.0% from among the
setting items using [ ► J and [

a

] keys. Then press

[SET] again to go into the calibration mode and

calculate and display zero point using a simulation
signal.

ADJUST READY

0

.

1

%

Pressing [ ► ], “ADJUST READY’" appears as

shown left and Ihe system goes into a calibration

ready condition. (Note 1)

N1:FL0Vr CAL OX

* CAL OX ADJ.

Pressing [SET], “*CALt)% ADJ.” appears as shown

left and the sy.siem starts the zero calibration. The

zero calibration takes about 3 to

6

seconds. (Note 2)

SET

N1:FLDi CAL OX

0

.

0

%

Xcw ly calibrated zero point appears.

SET

N l i F L O M C A L O X Pressing [SET], the system returns to the items

selection sequence.

Notes;

l*To cancel iicro c<tJibration, press [

a

], The system returns to the point where the zero point

display appears,

2,Calibration time depends on the cxcilation iroquoncy

(24 Uz:

3 sec, 12 IIz and

6

IIz;

6

sec).