6)groiindirif, Figure 4.10 grounding procedure, Toshiba – Toshiba LF494 User Manual

Page 23

Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

Advertising
background image

TOSHIBA

6 F 8 A 0 7 . 7 4

(6)Groiindirif<

The grounding terminal of the LF494/LK404 Q

o

a

meter should be grounded with lOUohm

or less ground resistance. Use a heavy copper braid or wire (cross-sectional area

5 , 5

nun"

minimum) to ground the lerminal and make it as short as possible. The terminal is M4 siiic
and an M4-siiie crimped ring fug should be used to connect the wire to the terminal. Avoid a

common ground where earth currcni may fliw. An independent ground is preferable, See

Figure 4.1(1.

To prevent a iwo-poini grounding, ground the shielded cable on the receiving instrument

side,

Figure 4.10 Grounding Procedure

- 2 2 -