10*7 Remote Zero Adjustment

On-slream zero adjustmciu in a

z e r o

flow rale condition can be started with an exiernal signal.

To do this, set D[ as a zero adjustment start signal. See R.2.9, “Digital I/O”

Start signal requirements:

Zero adjust men I

stari signal

{[ level

L level

! Zero

j

1C seconds ^ adjustment

start

20

seconds

The start signal must be set to H level firsi, then it must go to L level after Ihe passage of

more than

10

.seconds but not more than

20

seconds, as shown above. (If the signal does not

go to L level within liiis specified period, it will be ignored.) As soon as the signal goes to L

level, zero adjustment sequence starts.

-