2 output status for errors and alarms, Toshiba – Toshiba LF494 User Manual

Page 135

Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

Advertising
background image

TOSHIBA

6 F 8 A 0 7 7 4

12.2 Output Status for Errors and Alarms

The flowmeier datii display, current and puJsc outputs will become as follows if an error or

alarm occurs.

Krror or alarm
nuissage_____

ROM ERROR

(Note

1

)

RAM ERROR

PARAMETER FAIL

___________ (Note 2}

EX. CERR OPEN

EX.CURR ERROR

ADC. ERROR

EMPTY

Data

display

Zero

Zero

Zero

Zero

Zero

Cun^nt output
(4-20mA)

4mA

(Note 3}

Totali/cr and

pulse output

Stopped

Stopped

Stopped

Slopped

Stopped

Stnp[>cd

Stopped

Remarks

After piTwer-up,

no mcasurciticnl

starts,

Zero adjustment

(on-stream at zero

flow rate) cannot be

eomlucted.

Zero adjusimetit

{on-stream at ^cro
flow ratej cannot be
conducted.

INVALID TOTAL

Measured

data

Measured

data

Measured

data

The error message
disappears if yen
clear (reset) I he
totalizer.

H.ALARM

Measured

data

Measured

data

Measured

data

L.ALARM

Measured

data

Ml:,^a5urcd

data

Measured
data

Notes

1

.

2

.

3.

The display and output may not be as indicated depending on the nature of the ROM error.

if parameters related to the current output are defective, the current output may not be

exactly 4mA.

[ji case of the converter for special specifications (refer lo 7.3.3 ‘‘Converter for Special

Specifications’"),

select the value which he set 4-20mA Alarm Output. For detail, see See

e.2,ia""4-20mA Alarm Output ”,

-