(3) Flow rate ladication

la

not stable.

START

ÏS power supply voltage

within the specified range?

NO

YES

Are cables securely
connected?

YES

Is the flowmeter grounded

with 100 Q or less ground

resistance?

YES

Is the detector pipe filled
with fluid?

YES

Is the Huid carrying
bubbles?

NO

Is there high-voltage or large

current cable or equipment

near the flowmeter?

NO

YES

Contact your nearest Toshiba
representative.

6 F 8 A 0 7 7 4 .

Use a power supply within
the specified range.

NO

Connect each cable securely

to the terminal board.

Ground the flowmeter with a

copper braid or wlre(5.5 mm^

minimum) to a good earth

ground (100 Dor less ground

resistance).

NO

Design piping so that the

)

*

detector pipe is filled with

fluid all the time.

Design piping so that the tluid
does not carry bubbles.

YES

Install the flowmeter away

'

l

from equipment producing

}

electromagnetic or

electro- static interference.

"