c o n t i n u e d f r o m p r e v i o u s p a g e

Clear (reset) the totalizer.

Clear (reset) the totalizer by pressing [ ► ] key.

[f you have changed the counting rate, clcnr (reset) the totaiizer before you start the

totalizer

Start the totalizer.

■ Start the totalizer by pressing '

a

] key and make sure “C’' is shown on the display,

(Note 3)

Notes:

1. When the flowmeter has no optional digital input function (ninth column of Type

Specification Code; T). the movement of the totalizer and pulse output after power on is

decided by setting contents of optional digital input (Dl) and digital input detective level,

Necessary, set the DI function as “fh NO USt”.

To

lei

them start,

operate Che

control keys on the panel (Option) according to Totalizer

Operation

in

the

next subsection,

2.

It is preferable to set the indicating unit (UNIT 1 ancl'or UNIT 2) to one of Ihe units ap­

propriate for totalization just to make sure it is operating correctly,

3.

If (he indicating unit (UNIT 1 andi^or UNIT 2) is not the one for totalization, “C" docs not

appear on the display.

4.

Example for counting rate:

Example

Case Range ;36LM]m'Vh (Im’Vs)

Counting raie(palse rate)

Min.: 36[:'0(iTv'/h) / 3ri00n(]()(pulses/h) =0, 001 m^=1 I

Max.: 36fK)(m'Vh) / 3.fi(pulses/h) =100Q

-