8,2ЛЗ Empty Pipe AJarm

The empty pipe alarm is used to notify that the detector pipe is not filled with fluid, [f ни

empty pipe condition occurs,

a

message ♦ EMPTY appears. You Ciiii enable or disable this

function here.

In case of the converter for special specifications (refer to 7.3.^ "Converter for Special

specifications”), you can select Empty Pipe Alarm setting OFE, ONLEVELl, ON LHVBL2>

or ON LEVEL3.

Empty Pipe Alarm Functions

Empty pipe
alarm

Function

0: OFF

Disable empty pipe alarm

1: ON LEVEL!

Enable empty pipe alarm (detective level high)

2:

ON LEVELZ

Enable empty pipe alarm (detective level middle)

3: ON LEVELS

Enable cmply pipe alarm (detective level low)

NOTE: Setting enable empty pipe alarm, normally select ‘"1:0N LEVEL 1”.

In case of difficult condition to detect empty pipe alarm, select

-2 ON LEVEL2^'or”3:ON LEVELS '.

Proceed as follows to check or change the empty pipe alarm enablc/disablc status.

To check the empty pipe alarm enable/disable status:

Key operation

Display example

Description

SKI

J1: EMPTY ALM

ON

Press [SET] first to ,star£ the items selection sequence
and select Jl: EMPTY ALM from among the

configuration items using [ ► ] and [

a

. ] keys. TTien

pre,ss [SET] again to display the current empty pipe

Hlann enable/disable status.

SET

___ _____

Jl: EMPTY ALM

Pressing [SET], the system returns to the items selec­

tion sequence.

-

79

-