Counting Rate (pulse rate)

When ;hc tuliilizer is us&d for toial flow measurement, per-couut (pulse) value is the

couiUing rate. Pulse output is also available for external totalization. In this ilcrn, the

counting rate and the pulse width for pulse output can be checked or changed. The

counting rate is set using three digits and exponential quotient.

For example, 0.123

-> 1.23^-1

(3,23>i

10

'' m-)

Proceed as follows to check or change the counting rate.

■ To check the counting rate

and

pulse

width:

Key operation

Display example

Description

SliT

Gl: COUNT RATE

I. 00E-2m’

Press [SET] first to start the items selection sequence
and se 1 ect Cl: COUNT RATE from a mong t he
configuration items using [ ► ] and [

a

.

] keys. Then

press iSE’1’1 asain to display the current counting rate.

SET

Gl: COUNT RATE

Prcsssing [SETj, the system returns to the items
selection sequence.

Counting rate and pulse width can be selecicd by the configuration items as follows:

Counting rate

d ^ COUNT RATE

Pulse width

G2: PLS. WIDTH

NOTKS

1, The counting rale should be set so that its rale for 100?^ flow rate output is

within the range from to 36U()()(K) p u l s e I f you try to set the counting rale
outside of this range, an error message ^ H. OVER SPEC or i L. OVER SPEC

appears.

Set the counting rate within the specified range,

Example

Case Range :36tl0m’^/h (ImVs)

Counting rate(pulse rate)

Min.: 36OO(m‘Vh)/36UOOOD(pulses/h)=^0, 001 j

Max.: 3fiiH)(m^/h) / 3.6(pulses/h) =1000

-

71

-