6 adjustment, 1 adjust balance – KERN & SOHN DBS 60-3 User Manual

Page 31

Advertising
background image

DBS-BA-e-1311

31

6 Adjustment

6.1 Adjust balance

As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.

• Observe stable environmental conditions. A warming up time (see chapter

1) is required for stabilization.

• Carry out adjustment with placed sample dish. Ensure that no objects are

within the sample dish.

• Carry out adjustment as near as possible to the balance’s maximum

weight (recommended adjustment weight see chap. 1). Weights of
different nominal values (10g – 60g) may be used for adjustment but are
not optimal for technical measuring. Info about test weights can be found
on the Internet at: http://www.kern-sohn.com

 Press the Menu button to access the menu.

 Use navigation buttons   to select menu item

CAL”.

 Acknowledge using ENTER key, the current

setting is displayed. If necessary, use navigation
buttons

  to select “BAL”.

 Press ENTER to confirm and “WAIT” will be

displayed followed by the currently preset
adjustment weight.





Open heated cover

Advertising