KERN & SOHN DBS 60-3 User Manual

Page 52

Advertising
background image

52

DBS-BA-e-1311

4. Preset drying time for step 1

 Press ENTER to confirm. The currently preset

drying time for step 1 will be displayed.

 Use navigation buttons   to select the

required setting, 1 minute – 4 hours (1 minute
increments) can be selected

5. Set drying temperature for step 2

 Press ENTER to confirm and the currently

preset drying temperature for step 2 will flash
in the display.

 Use navigation buttons   to select the

required setting,
50°C – 200°C (1°C increments) can be
selected.

6. Set drying time or

M for step 2

 Acknowledge using ENTER key, the current

setting is displayed.

 Use navigation buttons   to select switch-

off criterion (TIME or

∆M) for step 2.

Setting

M

 Press ENTER to confirm selection. The

currently preset value for

∆M will be

displayed.

 Use navigation buttons   to select the

required setting, 0.01% – 0.1% (0.01%
increments) can be selected.

or

TIME setting

 Press ENTER to confirm selection. The

currently set drying time is displayed.

 Use navigation buttons   to select the

required setting, 0 minutes – 240 minutes
(1 minute increments) can be selected
If 0 minutes are entered, the next step will be
cancelled.

 Press ENTER to confirm.

Continue from step 8 for 2-step drying.

Continue from step 7 for 3-step drying.