6.2 Adjust temperature


We recommend sometimes to check the temperature value of the device using the
optional temperature calibrating set DBS-A01. The temperature is measured at two
points (100°C and 180°C) and can be adjusted at these two points.
The equipment should cool down first to room temperature after the last heating
period.

Preparation:
 Fix the temperature sensor on the

temperature calibration set in
accordance with the illustration.

 Remove sample dish and dish holder

from the moisture analyser.

 Install the temperature-calibration set

acc. to fig.

 Close heated cover.

If applicable, the error message “ol” will
be displayed, but can be ignored in this
case.

 On the temperature calibration set

switch-on the digital thermometer by the
ON/OFF button.