14 Instant help

Fault

Possible cause

Display is not lit up.

• The display unit is not switched on.

• The mains supply connection has been

interrupted (mains cable not plugged in/faulty).

• Power supply interrupted.

• Fuse has blown

The display does not change
when a sample is being loaded

• Sample dish / dish holder is fitted incorrectly.

The weight display changes
constantly / the stability display
 does not appear.

• Sample dish has contact with wind protection

device or heated cover.

• Draught/air movement

• Table/floor vibrations

• Electromagnetic fields / static charging (choose

different location/switch off interfering device if
possible)

Incorrect measuring result

• Check adjustment

• No resetting to zero before loading the sample

Measurement is taking too
long

• Incorrect setting shutoff criterion

Measurement is
not reproducible

• Sample is not homogenous

• Drying time is too short

• Drying temperature too high (e.g. oxidation

sample material, boiling point of sample
exceeded)

• Temperature sensor soiled or defective

Drying does not start

• Heated cover open

• The mains supply connection has been

interrupted (mains cable not plugged in/faulty).