KERN & SOHN DBS 60-3 User Manual

Page 50

Advertising
background image

50

DBS-BA-e-1311


 Press ENTER to confirm selection. The currently

present switch-off criterion will be displayed.

 Use navigation buttons   to select the

required setting,
TIME = time-controlled
∆M = automatic

TIME setting

 Press ENTER to confirm selection. The currently

set drying time is displayed.

 Use navigation buttons   to select the

required setting, 1 minute – 4 hours (1 minute
increments) or 4 hours - 12 hours (1 hour
increments) can be selected

 Press ENTER to save.

 Press ESC key to return to weighing mode. The

moisture analysis can be started (see chap. 8.2).
All the settings that have been made control the
drying process.

or

Setting

M

 Press ENTER to confirm selection. The currently

preset value for

∆M will be displayed.

 Use navigation buttons   to select the

required setting,
0.01% – 0.1% (0.01% increments) can be
selected.

 Press ENTER to save.

 Press ESC to return to Moisture analysis mode.

The equipment is ready to measure (see chap.
8.1). All the settings that have been made
control the drying process.