KERN & SOHN DBS 60-3 User Manual

Page 35

Advertising
background image

DBS-BA-e-1311

35


 The second temperature point to be tested will

be displayed.

 Confirm with the ENTER button, the second

heat-up phase is started.

 Temperature calibration for the second point

takes 15 min. Compare the displayed
temperature value on the thermometer with that
of the moisture analyser whilst the temperature
display is flashing. If the two values do not
match, correct them using the navigation
buttons

  and confirm with the ENTER

button Numerical input, see chap. 7.1.1.

 The adjustment is completed when “END” is

displayed. The device returns to the menu
automatically.
The adjustment protocol output is started upon
connection to an optional printer and activated
GLP function (see chap. 6.3).

 Press ESC to return to Moisture analysis mode.

• The adjustment protocol output is started upon connection to an

optional printer and activated GLP function (see chap. 6.3).