Call up menu:

 Press the Menu button to access the menu and

the first menu item “PRoGRM” will be
displayed.

 Use navigation buttons   to select menu

item “CAL”.

 Acknowledge using ENTER key, the current

setting is displayed. If necessary, use
navigation buttons

  to select “TEMP”.

 Press ENTER to confirm and the first

temperature point to be tested will be displayed.

Press START and the first heating-up period

will start.

• Do not leave the heated cover open for more than 1 minute during the

heating-up period, otherwise the error message “ERR.100” will appear.
In this case press ESC and restart menu item “TEMP”.

• Attention: some parts e.g. ventilation grilles and inspection windows

may become very hot during operation.

• The adjustment must be made within 15 minutes when the temperature

display flashes, otherwise the temperature adjustment will be aborted
(ABORT will be displayed).

 Temperature calibration for the first point takes

15 min. Compare the displayed temperature
value on the thermometer with that of the
moisture analyser whilst the temperature
display is flashing. If the two values do not
match, correct them using the navigation
buttons

  and confirm with the ENTER

button Numerical input, see chap. 7.1.1.