8 moisture analysis, 1 start drying – KERN & SOHN DBS 60-3 User Manual

Page 40

Advertising
background image

40

DBS-BA-e-1311

8 Moisture analysis

8.1 Start drying

First set the drying parameters as described in chap. 9 during initial start-up.

 Or set the drying parameters as described in chap. 9

or load from store a suitable drying program for the
sample, see chap. 9.1.
Ensure that the equipment is in Moisture analysis
mode. Press ESC if necessary.

 Open heated cover

 Deposit a sample dish acclimatised to room

temperature.

 When the heated cover is closed, wait for the stability

display [

] and press the TARE button to reset to

zero.

Load sample within 30 minutes otherwise the message “TIM.oUT” will be
displayed. In this case, press ESC and reset to zero.

 Place sample in sample dish.

Sample preparation see chap. 12.4
Ensure minimum initial weight > 0.02 g

Advertising