9 adjustment, 1 verified weighing systems – KERN & SOHN FA 6T-3M User Manual

Page 15

Advertising
background image

KFB/KFN-TM-BA_IA-e-1220

15

6.9 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,

each display unit with connected weighing plate must be coordinated - in compliance

with the underlying physical weighing principle - to the existing acceleration due to

gravity at its place of location (only if the weighing system has not already been

adjusted to the location in the factory). This adjustment process must be carried out

for the first commissioning, after each change of location as well as in case of

fluctuating environment temperature. To receive accurate measuring values it is also

recommended to adjust the display unit periodically in weighing operation.

In weighing systems with a resolution of < 15 000 dividing steps an

adjustment is recommended.

In weighing systems with a resolution of > 15 000 dividing steps a

linearisation is recommended (see chap. 6.10).

Prepare the required adjustment weight. The weight to be used depends

on the capacity of the scale. Carry out adjustment as near as possible to

the scale’s maximum weight. Info about test weights can be found on

the Internet at: http://www.kern-sohn.com.

Observe stable environmental conditions. Stabilisation requires a certain

warm-up time.

6.9.1 Verified weighing systems

In verified weighing systems the menu item for adjustment „P2 mode“ is

blocked.
KERN KFB-TM
To disable the access lock, destroy the seal and actuate the adjustment

switch. Position of the adjustment switch see chap. 6.11
KERN KFN-TM

To override the blocked access you will have to destroy the seal before calling

up the menu and to short-circuit the two contacts on the circuit board [K2],

using a jumper (See chap. 6.11).

Attention:

After destruction of the seal the weighing system must be re-verified by an

authorised agency and a new verification wire/seal mark fitted before it can be

reused for applications subject to verification.

Advertising