11 Instant help

In case of an error in the program process, briefly turn off the display unit and

disconnect from power supply. The weighing process must then be restarted from the

beginning.

Help:
Fault

Possible cause

The displayed weight does

not glow.

The display unit is not switched on.

Mains power supply interrupted (mains cable

defective).

Power supply interrupted.

(Rechargeable) batteries are inserted incorrectly or

empty

No (rechargeable) batteries inserted.

The displayed weight is

permanently changing

Draught/air movement

Table/floor vibrations

Weighing pan has contact with other objects.

Electromagnetic fields / static charging (choose

different location/switch off interfering device if

possible)

The weighing result is

obviously incorrect

The display of the balance is not at zero

Adjustment is no longer correct.

Great fluctuations in temperature.

Warm-up time was ignored.

Electromagnetic fields / static charging (choose

different location/switch off interfering device if

possible)

Should other error messages occur, switch display unit off and then on again. If the

error message remains inform manufacturer.