Annex page 2 of 7

Annex to Test Certificate No. DK0199-R76-11.04

Issued by DELTA

Indication of zero

Semi-automatic subtractive tare

Acting upon significant fault

Weighing unstable samples

Real time clock (optional)

3.

Technical data

3.1

Indicator

Type

KFN-TM / KFB-TM

Accuracy class

III or IIII

Weighing range

Single-interval, multi-interval or multi-range

Maximum number of verification scale intervals (n) 6000 for single-interval

2

×3000 for multi-interval and multi-range,

however limited to 1000 for Class IIII

Minimum input voltage per VSI

1 µV

Maximum capacity of interval or range (Max

i

): n

i

× e

i

Verification scale interval, e

i

=

Max

i

/ n

i

Initial zero-setting range:

± 10 % of Max

Maximum tare effect:

100 % of Max

Fractional factor (pi)

0.5

Excitation voltage

5 VDC

Circuit for remote sense

Active, (see below)

Minimum input impedance

87 ohm

Maximum input impedance

1600 ohm

Connecting cable to load cell(s):

See Section 3.1.1

Supply voltage:

9 - 12 VDC
230 VAC using external Vac/2Vdc adapter

Operating temperature range

Min / Max = -10 °C / +40 °C

Peripheral interface(s)

See Section 4

3.1.1

Connecting cable between the indicator and the junction box for load cell(s),

if any

3.1.1.1 4-wire system

Line

4 wires, shielded

Maximum length

The certified length of the load cell cable, which shall be con-

nected directly to the indicator.

3.1.1.2 6-wire system

Line 6

wires,

shielded

Maximum length

227 m/mm

2

Maximum resistance per wire

3.8 ohm