2 non verifiable weighing systems – KERN & SOHN FA 6T-3M User Manual

Page 18

Advertising
background image

18

KFB/KFN-TM-BA_IA-e-1220

6.9.2 Non verifiable weighing systems

Call up menu:

1. Switch-on balance and during the selftest press

.

2. Press subsequently

,

,

the first menu block

„PO CHK“ will be displayed.

3. Press

repeatedly until „P3 CAL“ will be displayed.

4. Confirm with

; press

repeatedly until „CAL“

appears.

5. Acknowledge using

, the current setting is displayed.

Press

to confirm; press

to select setting.

noLin = adjustment

LineAr = linearization, see chap. 6.10

How to carry out adjustment:

Confirm menu setting „noLin“ by

.

Ensure that there are no objects on the weighing plate.

Wait for stability display, then press

.

The currently set adjustment weight will be displayed.

To change by using the navigation buttons (see chap. 2.1.1)

select the desired setting, the active digit is flashing.

Acknowledge with

.

Carefully place adjusting weight in the centre of the weighing

plate. Wait for stability display, then press

.

After the adjustment the balance will carry out a self-test.

Remove adjusting weight during selftest, balance will return

into weighing mode automatically. An adjusting error or

incorrect adjusting weight will be indicated by the error

message; repeat adjustment procedure.

Advertising