KERN & SOHN FA 6T-3M User Manual

Page 22

Advertising
background image

22

KFB/KFN-TM-BA_IA-e-1220

Notes on verified weighing systems

KFB-TM:

Access to conductor plate:

Remove seal

Open display unit

The application of the display unit as a weighing system able to be verified

requires that the contacts of the circuit board are short-circuited with the help

of a jumper [K1].

For non verifiable weighing systems remove the jumper.

In verified weighing systems the menu item for adjustment, „P2 mode“ will be

blocked.
To disable the access lock, destroy the seal and actuate the adjustment switch.

Position of seals and adjusting switch

1. Self-destroying seal mark

2. Adjustment switch

3. Cover of adjustment switch

4. Self-destroying seal mark

[K1]

Advertising