2 non-verified weighing systems – KERN & SOHN FA 6T-3M User Manual

Page 20

Advertising
background image

20

KFB/KFN-TM-BA_IA-e-1220

After linearisation the balance will carry out a self-test.

Remove adjusting weight during selftest, balance will return

into weighing mode automatically.

6.10.2

Non-verified weighing systems

Call-up menu item P3 CALCalLiner, see chap. 6.9.1

Confirm by

, the password query „Pn“ will be displayed.

Press

,

,

subsequently.

Ensure that there are no objects on the weighing pan.

Wait for stability display, then press

.

When “Ld 1“ is displayed, put the first adjustment weight (1/3

max) carefully in the centre of the weighing platform. Wait for

stability display, then press

.

When “Ld 2“ is displayed, put the second adjustment weight

(2/3 max) carefully in the centre of the weighing platform.

Wait for stability display, then press

.

When “Ld 3“ is displayed, put the third adjustment weight

(max) carefully in the centre of the weighing platform. Wait

for stability display, then press

.

After a successful linearisation the balance will carry out a

self-test. Remove adjusting weight during selftest, balance

will return into weighing mode automatically.