9 service, maintenance, disposal, 1 clean, 2 service, maintenance – KERN & SOHN FA 6T-3M User Manual

Page 43: 3 disposal, 4 error messages

Advertising
background image

KFB/KFN-TM-BA_IA-e-1220

43

9 Service, maintenance, disposal

9.1 Clean

Before cleaning, disconnect the appliance from the operating voltage.

Do not use aggressive detergents (solvents or similar).

9.2 Service, maintenance
The appliance may only be opened by trained service technicians who are authorized

by KERN.

Before opening, disconnect from power supply.

9.3 Disposal
Disposal of packaging and appliance must be carried out by operator according to

valid national or regional law of the location where the appliance is used.

9.4 Error messages

Error

message

Description

Possible causes

- - - - -

Maximum load exceeded

Unload weighing system or reduce

preload.

- - ol - -

Err 1

Incorrect data input

Follow format “yy:mm:dd“

Err 2

Incorrect time entry

Follow format “hh:mm:ss“

Err 4

Zeroing range exceeded due

to switching-on balance or

pressing

(normally 4%

max)

Object on the weighing plate

Overload when zeroing

Err 5

Keyboard error

Err 6

Value outside the A/D changer

range

Weighing plate not installed

Damaged weighing cell

Damaged electronics

Err 9

Stability display does not

appear

Check the environmental

conditions.

Advertising