12 Installing display unit / weighing bridge

Installation / configuration of a weighing system must be carried out

by a well acquainted specialist with the workings of weighing

balances.

12.1 Technical data

Supply voltage:

5 V/150mA

Max. signal voltage

0-10 mV

Zeroing range

0-2 mV

Sensitivity

2-3 mV/V

Resistance parameter

80 - 100 Ω, max 4 items per 350 Ω load cell

12.2 Weighing system design
The display unit is suitable for connection to any analogue platform in compliance

with the required specifications.

The following data must be established before selecting a weighing cell:

Weighing balance capacity

This usually corresponds to the heaviest load to be weighed.

Preload

This corresponds to the total weight of all parts that are to be placed on the

weighing cell such as upper part of platform, weighing pan etc.

Total zero setting range

This is composed of the start-up zero setting range (± 2%) and the zero

setting range available to the user via the ZERO-key (2%). The total zero

setting range equals therefore 4 % of the scale’s capacity.

The addition of weighing scales capacity, preload and the total zero setting

range give the required capacity for the weighing cell.

To avoid overloading of the weighing cell, include an additional safety

margin.

Smallest desired display division

Verifiability, if required

The application of the display unit as a verified weighing system requires

that you short-circuit the two contacts [K1] of the circuit board, using a

jumper; for position see chap. 6.11.

Remove the jumper for weighing systems not able to be verified.