12.4 Configure display unit

12.4.1

Verified weighing systems
(contacts of circuit board [K1] short-circuited by means of jumper)

For menu overview see chap. 8.2.
In verified weighing systems the menu item for calibration „P2 mode“ is blocked.

KERN KFB-TM:
To disable the access lock, destroy the seal and actuate the adjustment switch.

Position of the adjustment switch see chap. 6.11
KERN KFN-TM

:

To override the blocked access you will have to destroy the seal before calling up the

menu and to short-circuit the two contacts on the circuit board [K2], using a jumper

(See chap. 6.11).
Attention:

After destruction of the seal the weighing system must be re-verified by an

authorised agency and a new verification wire/seal mark fitted before it can be

reused for applications subject to verification.

Call up menu:

Switch-on balance and during the selftest press

.

Press

,

,

subsequently, the first menu block

„PO CHK“ will be displayed.

Press

repeatedly until „P2 mode“ will be displayed.

Operate the adjustment switch (models KFB-TM).

Press

and use

to select the weighing scales type.

Single-range balance

Dual range balance

Multi-interval balance

Advertising