Real time clock

If it is available in the instrument, the real time clock can be activated to get printout with day and
time information.

2.2.9

Operator information messages

The weight indicator has a number of general and diagnostic messages which are described in detail in
the user’s guide.

2.2.10 Software version

The software revision level is displayed during the power-up sequence of the instrument.

The approved software version is 1.07 and 1.08.
The software version 1.08 includes possibility of multi-range.

2.2.11 Totalisation

The indicator can be configured with a totalisation function, adding actual weight display values to the
memory when pressing “M+” key if the equilibrium is stable.

Pressing “MR” key displays the total accumulated weight.
Pressing “M+” and “MR” key will clear the totalised value.

2.2.12 Battery operation

The indicator can be operated from an internal rechargeable battery, if this option is installed.

3.

Technical data

The KFN-TM / KFB-TM / BFB / IFB / SFB / UFB / UFN / NFB / BFN / NFN weighing instruments
are composed of separate modules, which are set out as follows:

3.1

Indicator

The indicators have the following characteristics:

Type:

KFN-TM / KFB-TM / BFB / IFB / SFB / UFB / UFN / NFB
/ BFN / NFN

Accuracy class:

III and IIII

Weighing range:

Single-interval, multi-range (2 ranges) or multi-interval (2
partial intervals)

Maximum number of Verification
Scale Intervals:

≤ 6000 (class III), ≤ 1000 (class IIII) for single-interval

≤ 3000 (class III), ≤ 1000 (class IIII) for multi-range and
multi-interval

Maximum tare effect:

-Max within display limits

Fractional factor:

p'i = 0.5

Minimum input voltage per VSI:

1

µV

Excitation voltage:

5 VDC

Circuit for remote sense:

present on the model with 7-terminal connector

Minimum input impedance:

87 ohm

Maximum input impedance:

1600 ohm

Mains power supply:

9 - 12 VDC / 230 VAC, 50 Hz using external adapter

Operational temperature:

-10 °C to +40 °C