Section 8 Device Message Details

8.3.5 CALCulate Subsystem

The CALCulate subsystem performs the performance measurement setting and measured result display.

: CALCulate

; DATA^

<string>

: CALCulate : DATA? <string>

Command type

Sequential command

Parameter

<string> = <STRING PROGRAM DATA>

Response

<string> = <STRING RESPONSE DATA>

See the Table 8-1 for <string>.

Function

Reads the measured results.

Example of use

> : CALCulate : DATA? "CURRent : JAMPlitude : PTPeak"

< 1.00

Table 8-1 String

Item

<string>

Response Format

Jitter manual measurement

CURRent

Peak to Peak

"[CLiRRent:] JAMPlitude : PTPeak"

Form 6

4-Peak

"[CURRent:! JAMPlitude : PPEak"

Form 6

-Peak

"[CURRent:] JAMPlitude : MPEak"

Form 6

RMS

"[CURRent:] JAMPlitude : RMS"

Form 6

LAST

Peak to Peak

"LAST : JAMPlitude : FfPeak"

Form 6

-rPeak

"LAST : JAMPlitude : PPEak"

Form 6

-Peak

"LAST : JAMPlitude : MPEak"

Form 6

RMS

"LAST : JAMPlitude : RMS"

Form 6

8-24

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com