Section 8 Device Message Details
:
SOURce : TELecom
: OFFSet <numeric>
Command type
Sequential command
Parameter
<numeric> = <DECIMAL NUMERIC PROGRAM DATA>
-50 to +50 ... step 0.!
Function
Performs the frequency offset setting.
Example of use
Sets the value of frequency offset at +25.
> : SOURce ; TELecom : OFFSet 25
: SOURce : TELecom : OFFSet?
Command type
Response
Function
Example of use
Sequential command
<NR1 NUMERIC RESPONSE DATA>
-50 to +50 ... step 0.1
Makes a query of the value of frequency offset.
> : SOURce ; TELecom : OFFSet?
<-25
: SOURce ; JITTer : RANGe <range>
Command type
Parameter
Function
Restriction
Example of use
Sequential command
<range> = <CHARACTER PROGRAM DATA>
U13200......... 3,200
UI
pp
UI1600 ......... 1,600
UI
pp
UI800............ 800
UI
pp
UI80 ............. 80
UI
pp
UI40 ............. 40 UI
pp
UI20 ............. 20
UI
pp
UI05 ............. 0.5
UI
pp
Performs setting of the transmission jitter generation range.
Effective when transmission jitter is generated.
Restrictions on setting apply depending on the Bit Rate.
Bit Rate
Ul Range
lOG
3,200, 80, 0.5 UI
5G
1,600, 40, 0.5 UI
2.5 G
800, 20, 0.5 UI
Sets the transmission jitter generation range at 0.5
UI
pp
.
> : SOURce : JITTer ; RANGe U105