5 reading, writing and clearing status registers, 1) reading and writing methods – Anritsu MP1777A User Manual

Page 69

Attention! The text in this document has been recognized automatically. To view the original document, you can use the "Original mode".

Advertising
background image

6.5

Reading, Writing and Clearing Status Registers

(1) Reading and writing methods

The table below shows the methods of reading and writing the details of status registers.

Table 6-7 Reading and Writing Methods of Status Registers

Section 6 Status Report

Register

Reading Methods

Writing Methods

Status byte register

Serial poll

Seven-bit status byte and RQS message

bit are returned. In this case, the status

byte value does not change.

Common *STB? query

The details of the status byte register

and one numerical value from MSS

(Master Summary Status) are returned.

Disabled

Service request enable register

Common *SRE? query

Common *SRE? command

Standard event status register

Common *ESR? query

In this case, the details of the register is

cleared after it is read.

Disabled

Standard event status enable

register

Common *ESE? query

In this case, the detail of the register

will not change.

Common *ESE? command

SCPI event register

SCPI command (STATus subsystem)

; STATUS : ... : EVENt?

In this case, the details of the register

will be cleared.

Disabled

SCPI enable register

SCPI command (STATus subsystem)

; STATUS ; ... ; ENABle

In this case, the details of the register

will not change.

SCPI command (STATus subsystem)

: STATUS : ... : ENABle?

SCPI Transition filter

SCPI command (STATus subsystem)

: STATus : ... ; PTRansition?

: STATus : ... ; NTRansition?

In this case, the details of the register

will not change.

SCPI command (STATus subsystem)

: STATus : ... : PTRansition

: STATus ; ... : NTRansition

Error/Event Queues

SCPI command

: SYSTem ; ERRor?

Disabled

NOTES:

1

.

The SCPI event register, SCPI enable register and SCPI Transition filter correspond to the event register,

enable register and Transition filter in the status registers specified by SCPI and the device-specific status

registers.

2. See the Section 5 "Common 1EEE488.2 Commands" for more information on the common commands and

queries and the Section 8 "Device Message Details” for more infonnation on the SCPI commands.

6-8

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com