8.3 Device-Specific Commands
: STATUS : OPERation : PTRansition <numeric>
Command type
Sequential command
Parameter
<numeric> = <DECIMAL NUMERIC PROGRAM DATA>
0 to 32,767 ......... Sum (decimal digit) of transition filter bits
Function
Sets the transition filter (a change in the positive direction) of the OPERation status register.
Set at 1.
Example of use
Sets the bit being measured at PTRansition.
> : STATUS
; OPERation : PTRansition 16
:
STATUS
:
OPERation
:
PTRansition?
Command type
Sequential command
Response
<numenc> = <NR1 NUMERIC RESPONSE DATA>
0 to 32,767 ......... Sum (decimal digit) of transition filter bits
Function
Makes a query of the details of the transition filter (a change in the positive direction) of the
OPERation status register.
—
Example of use
When the bit being measured is set at PTRansition.
> ; STATUS : OPERation ; PTRansition?
< 16
:
STATUS
:
OPERation
:
NTRansition <numeric>
Command type
Sequential command
Parameter
<numeric> = <DECIMAL NUMERIC PROGRAM DATA>
0 to 32,767 ......... Sum (decimal digit) of transition filter bits
Function
Sets the transition filter (a change in the negative direction) of the OPERation status register.
Set at 1.
Example of use
Sets the INSTrument status register summary at NTRansition.
> : ST.ATus : OPERation : NTRansition 8192