8.3 Device-Specific Commands

8.3.4 Display Subsystem

The Display subsystem performs control (setting and display) of screens.

Display

: DSELect

:SETUP

: RESult

n

[:NAME]

[;NAME]

[;NAME]

:NAME]

; JITTer

MODE -

MODE?

UNIT -

UNIT?

<display>

<display>

<time>

<mode>

<unit>

; Display : DSELect [:NAME] <display>

Command type

Sequential command

Parameter

<display> = <STRING PROGRAM DATA>

"SETup"....... Setup screen

"TMENu" .... Test menu screen

"RESult" Result screen

"T&R".......... Test menu & Result screen

Function

Selects the screen dividing method.

Example of use

Selects the split screen display of the Test menu screen and Result screen.

> : Display : DSELect : MANE "T&R" or : DISPlay : DSELect "T&R"

: Display : DSELect [:NAME]?

Command type

Sequential command

Response

<display> = <STRING RESPONSE DATA>

"SET" .......... Setup screen

"TMEN"........Test menu screen

"RES"............Result screen

"T&R"...........Test menu & Result screen

Function

Makes a query of the screen dividing status.

Example of use

> : DISPlay : DSELect : NAME? or : DISPlay : DSELect?

< "T&R"