A.1 Command Errors

Table A-1 Error Messages of Command Errors

Code

Message

Error Detection Condition

-101

Invalid character

An invalid character i.s included in the header or the parameter.

E x a m p le :

1. A # is included in the header.

: SENSe : TELecom : #RATe Ml39

2. A $ is included in the parameter.

: SENSe : TELecom : PATTem : TYPE PRB$11

-104

Data type error

The parameter type is different to the type specified.

-105

GET not allowed

A Group Execute Trigger was sent to the program message.

-108

Parameter not allowed

The number of parameters is greater than the number specified.

E x a m p le :

Four parameter have been sent to the Section where two param­

eters were originally specified.

; SENSe : TELecom : DEMux : ROUTe M34,1,M8,2

-112

Program mnemonic too long

A program mnemonic comes in 12 or more characters.

E x a m p le : A

colon is missing from the header.

: CALCulate : TELecom : PERFormanceTYPE G821

-113

Undefined header

Although the header syntax is correct, it is not defined by the device.

E x a m p le :

The header was erroneously written.

; SENSe : TELecom ; DEMux : ROOT M34.1

-120

Numeric data error

An error was found in the numeric data.

-121

Invalid character in number

An inappropriate character is included in the numeric data.

E x a m p le :

A sign is included in the numeric data.

: SENSe : TELecom : M2 : MCHannel 3$

-130

Suffix error

An error was found in the suffix.

E x a m p le :

A part of data was set in alphabet while the parameter is numeric

data.

: SENSe ; TELecom ; M2 ; MCHannel 31

-144

Character data too long

The character data is more than 12 characters.

-150

String data error

The string data comes outside the specified range.

E x a m p le :

Single and double quotations are mixed.

: Display : MENU ; NAME "SYSTem"