6.5 Reading, Writing and Clearing Status Registers

(2) Clearing and resetting methods

The table below shows clearing and resetting methods of status registers and the range of their effects.

Table 6-8 Clearing and Resetting Methods of Status Registers

Register

*RST

*CLS

Power

Source ON

#

STATus:PRESet

Other Clearing

Methods

Status byte register

No change

Clear

Clear

No change

Service request

No change

No change

Clear

No change

Execution of *SRE 0

Standard event status

register

No change

Clear

Clear

No change

Reading of events by

*ESR? to clear the status

register

Standard event status

enable register

No change

No change

Clear

No change

Execution of *ESE 0

SCPI event register

No change

Clear

Clear

No change

Reading of events by .

: STATUS:...: EVENC to

clear the status register

1

SCPI enable register ! No change

No change

Reset

Reset

Execution of

: STATUS:... : ENABle 0

Execution of the

SCPI Transition filter

No change

No change

Reset

Reset

: STATUS:...: PTRansition 0

: STATUS:...: NTRansition 0

Error/event queue

No change

Clear

Clear

No change

Reading of all the events

by : SYSTem : ERRor?

NOTES :

The SCPI event register, SCPI enable register and SCPI transition filter correspond to the event register, enable

register and transition filter in the status registers specified by SCPI and device-specific status registers.

Cleared (or reset) when the power source is switched ON with the *PSC (Power-ON Status Clear) flag set as

true by the common PSC command.

The table below shows the reset values of registers influenced by the : STATus : PRESet command.

Register 0

Enable/Filter

Reset Value

OPERational status register

Enable register

ALLO

QUEStionable status register

PTRansition filter

ALL 1

NTRansition filter

ALLO

Enable register

ALL 1

INSTrument status register

PTRansition filter

ALL 1

NTRansition filter

ALLO

Enable register

ALL 1

Other status registers

PTRansition filter

ALL 1

NTRansition filter

ALL 1

6-9

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com