Martin Christ Epsilon 2-10D LSCplus User Manual

Page 106

Advertising
background image

Freeze-dryer EPSILON 2-10D LSCplus

10 Technical data

106

Version 04/2013, Rev. 1.5 of 16/12/2014 • sb

Translation of the original operating manual

Connection requirements

Electrical connection:

3 x 400 V / 50 Hz
(other supply data on request)

Protection class:

I

IP protection categoryaccording to DIN 60529:

11

Nominal power
- with vacuum pump Pfeiffer DUO 10 (3 ph):


6.8 kVA

Nominal current:

12.0 A

Mains fuse:

16.0 A

Equipment connections

Vacuum connection:

Stainless steel pipe, 20 mm

Media drain valve (standard):

Hose nozzle, 10 mm

Special equipment: media drain valve at the connector
block:

R3/4"

Aeration valve:

Hose nozzle, 6 mm

Air injection valve:

Hose nozzle, 6 mm

Mains input:

Cekon CEE (16 A 3ph)

Data interface (LAN)

RJ 45

Special equipment: water cooling system

Part number:

110827

Water feed flow temperature:

+15°C to +20°C

Water return flow temperature:

25°C max.

Water temperature difference:

10K max. with +15°C feed flow temperature
5K max. with +20°C feed flow temperature

Cooling water pressure:

2 bar min.
10 bar max.

p min. 1.5bar

Volumetric flow rate:

0.7 m

3

/h ( 11.7 l/min) max.

Heat dissipated by cooling water:

5,600 W

Cooling water feed flow connection:
Attention! When the unit is switched off, the water feed
flow must be shut off by the customer!

R3/4" with hose nozzle DN 13

Cooling water return flow connection:

R3/4" with hose nozzle DN 13

Option: cooling water flow controller (factory setting):

12 bar (refrigeration unit, high pressure side)