Freeze-dryer EPSILON 2-10D LSCplus

6 Operation

48

Version 04/2013, Rev. 1.5 of 16/12/2014 • sb

Translation of the original operating manual

Option: USB process recording (see chapter 6.6 - "Optional extensions and
special equipment")
• Select the "Process recording" function in the dialog box "Tools".
• Select the input fields ("Batch data"). A keyboard for the data input will

be displayed.

• If necessary, select the “Options” tab, choose between manual or

automatic recording, and define a recording interval.

• Press the "return" button in order to close the dialog box.
The process recording will now run in the background.

Fig. 28: Dialog box “USB process recording”


Option: Pressure increase test (only with the double-chamber method)
The pressure increase test can only be performed when the freeze-dryer
is equipped with an intermediate valve. The performance is possible in the
manual mode as well as in the program mode, Additionally, the pressure
increase test can be automatically performed as part of a program (see
chapter 6.5.3.1 - "Creating a program").
• Select "Pressure increase test" in the "Tools" dialog box.
• Enter the set values for the duration and maximum pressure increase

with the aid of the buttons.

• Start the pressure increase test. The test time will be displayed. After

the end of the test, a status message (pressure increase was
successful or failed) will be displayed together with the measured
values: