Freeze-dryer EPSILON 2-10D LSCplus

8 Maintenance and service

Version 04/2013, Rev. 1.5 of 16/12/2014 • sb

95

Translation of the original operating manual

Pos: 206 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 80 Wa rtun g u nd In stan dhalt ung/ 080 -00 10- 001 0-0 010 Eiskon dens ato rkam me r (a uß er Epsil on 2 -4) @ 2 5\m od_ 1404 984 098 898 _68. docx @ 18 434 9 @ 4 @ 1

8.1.1.1 Ice condenser chamber

Before each start-up, ensure that the ice condenser chamber is free from
water residues.
• Open the media drain valve to drain off any liquid. Then, close the

valve.

• If necessary, wipe the ice condenser chamber dry with a cloth.

Pos: 207 /01 0 Univ ersal mod ule/L eerz eile @ 0\m od_ 120 211 624 450 0_0. docx @ 11 4 @ @ 1

Pos: 208 /20 0 Ch rist/3 61 G T-BA L abo r-Pilot (PROJEKTE)/Epsilo n Sta nda rd/0 80 Wa rtu ng u nd I nstan dhal tung /08 0-0 010 -00 10- 002 1 Son der ausst attu ng: Tüv erschl ussrie gel aus Edels tahl Epsilon @ 25 \mo d_1 404 9846 771 36_ 68.d ocx @ 185 889 @ 4 @ 1

8.1.1.2

Special equipment: Stainless steel door latch at the loading door

Systems that are equipped with a stainless-steel or acrylic glass door with a
stainless-steel door latch have a slide pad on the door latch. This slide pad
is included in the scope of supply as a spare part. It must be inspected
regularly for signs of wear. When its thickness is < 0.5 mm, it must be
replaced.



1

Door latch

2

Slide pad

(part no. 177270)

3

Torx screws

Fig. 79: Door latch and slide pad


Replacing the slide pad:
Unscrew the two screws of the slide pad with a Torx

®

TX10 screwdriver,

remove the old slide pad, and install the new one.