2 insufficient vacuum – Martin Christ Epsilon 2-10D LSCplus User Manual

Page 90

Advertising
background image

Freeze-dryer EPSILON 2-10D LSCplus

7 Malfunctions and error correction

90

Version 04/2013, Rev. 1.5 of 16/12/2014 • sb

Translation of the original operating manual

Pos: 181 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 70 Stö run gen und Fe hlers uche /07 0-0 010 -00 20 Kei n au sreic hend es Vak uum @ 2 5\mo d_1 404 984 060 039 _68. docx @ 18 416 7 @ 3 @ 1

7.1.2 Insufficient vacuum

NOTE

The vacuum checks must be carried out when the ice condenser is frozen.

Pos: 182 /01 0 Univ ersal mod ule/L eerz eile @ 0\m od_ 120 211 624 450 0_0. docx @ 11 4 @ @ 1

Pos: 183 /20 0 Ch rist/3 60 G T-BA L abo r-Pilot (STANDARDMODU LE)/0 70 Stö run gen und Fe hlers uche /07 0-0 010 -00 20- 0010 Kleinfl anschv er bindu nge n @ 25\m od_ 140 498 406 108 6_68 .docx @ 1 841 81 @ 4 @ 1

7.1.2.1 Small flange connections

Leakages are often due to improper small flange connections between the
various components and hose connections or to leakages in the valves.
• Loosen the connection and place the centring ring (with sealing ring

inside) in a centred manner between the flange connections.

• Seal the connection with the clamping ring by tightening the wing nut.
• Ensure that the centring ring neither slips out of place nor gets jammed.

Fig. 75: Small flange and centring ring

Fig. 76: Small flange with centring ring
and small flange

Fig. 77: Attaching the clamping ring

Fig. 78: Tightened clamping ring