Freeze-dryer EPSILON 2-10D LSCplus

6 Operation

52

Version 04/2013, Rev. 1.5 of 16/12/2014 • sb

Translation of the original operating manual

Dialog box "Operating mode: select/start" (8)
After the set values have been entered for the process, the process can be
started manually with this function (see chapter 6.5.2 - "Manual freeze-
drying").

Fig. 34: Dialog box "Operating mode: select/start" (The version of the dialog box that is

displayed depends on the equipment of the freeze-dryer.)



Button "Unit Test"
Apart from the process phases of the freeze-drying process ("Freezing",
"Warm-up", "Main drying", and "Final drying") and the operating mode
"Defrosting", the button "Unit Test" is also available. This button opens a
selection of various tests with fixed parameters. After consultation with the
manufacturer, these tests can be performed in order to check the
functionality and processes of the freeze-dryer.

NOTE

Some of the available tests are also provided in the form of program
templates (see chapter 6.5.3.1 - "Creating a program").



Button "Stop" (9)
Pressing this button stops the current process. The system switches to the
standby status.