APPENDIX III

III.

RS-485 Half Duplex / Full (RS-422) Connection To 177A

RS-485 is a specialized interface that is very common in the

data acquisition world

.

An RS-485 network can be

connected in a 2 or 4-wire mode (bi-directional - half duplex - multi-drop Communications over a single or dual
twisted pair cable). All the devices in the same network should be connected in daisy chain (see diagram below).
Maximum cable length can be as much as 4000 feet.



















Tools and procedures to setup the 177A units for RS-485 communication:

Station ID, RS-485 and Terminator setup.

RS-232 to RS-485 Converter.

Cable


Station ID, RS-485 and Terminator setup: Each 177A on the same RS-485 bus must have its own unique ID (1-
255). This can be set via ME177A software or from the front panel Setup. Baud rate is 9600 by default.

Setup RS-485 from the 177A front panel:

Press Key

Display(blinking)

Display Description

Setup

Lo

Front panel Lock

Up

SS

System Setup

Right

bAu

Baud rate

Down

IT

Interface Type

Right

+232

RS-232

Down

485 - HAF

RS-485 Half duplex

Enter


Setup Terminator (continue from above):

Press Key

Display(blinking)

Display Description

Left

+485

Down

TEr

Terminator

Right

on

Press up/down to select terminator on / oFF

Enter

RS-485 Half Duplex

RS-232 or USB
to RS-485
Converter

Pin 2 3

177A

Unit 1

Termination enables

Termination enables

Up to 4000 feet

485 Data+

485 Data-

T+
R+
R-
T-

Pin 2 3

177A

Unit 2

Pin 2 3

177A

Unit n