Monroe Electronics Electrostatic Fieldmeter - Static Monitor - model 177A User Manual

Page 49

Advertising
background image

48

177A reply to read coil:

Unit# command

# of data

Coil Value

crcH

crcL

1-255 0x01

1

0=Off, 0xff=On

Write Channel LEDs or Relays (write coil):

From Master (PC software or PLC):
Unit# Command

CoilAdrH

CoilAdrL

New Value

0

crcH

crcL

1-255 0x05

0

0-24

0=Off, 0xff=On

The request consisted of the Unit address and the command code 05 followed by a 2 byte starting
address, specifying the variable to be set, a data byte, specifying the binary value to set ( ON = 0xff,
OFF = 0 ) and an additional byte always set to 0.

177A reply to write single coil is identical to the received packet:
Unit# Command

CoilAdrH

CoilAdrL

New Value

0

crcH

crcL

1-255 0x05

0

0-24

0=Off, 0xff=On

Read Channel display (read register):

From PC or PLC:
Unit# Command

RegAdrH

RegAdrL

LenH LenL crcH crcL

1-255 0x03

0

0-3

0

1

RegAdrL is as follows:

0

Channel 1

1

Channel 2

2

Channel 3

3

Channel 4

177A reply:

Unit# Command

# of data

dataH dataL crcH crcL

1-255 0x03

1

##

##

Write Channel display (write register):

From PC or PLC:

Unit# Command

RegAdrH

RegAdrL

dataH dataL crcH crcL

1-255 0x06

0

0-3

##

##

177A reply to write single Register is identical to the received packet:

Unit# Command

RegAdrH

RegAdrL

dataH dataL crcH crcL

1-255 0x06

0

0-3

##

##

Note:

If the Unit# is 0, the 177A units will execute the Write-Coil and Write-Register commands, but No reply.