APPENDIX IV

Communication with ME 177A using MODBUS Protocol

Read Channel LEDs or Relays (read coil)
Write Channel LEDs or Relays (write coil)
Read Channel display (read register)
Write Channel display (write register)

Read Channel LEDs or Relays (read-coil):

From Master (PC software or PLC):
Unit# Command CoilAdrH CoilAdrL lenHigh lenLow crcH

crcL

1-255 0x01 0

0-24

0

1

The request consisted of the unit# and the command code 01 followed by a 2 byte starting address,
specifying the variable to be read.

Length high must be 0, length low is 1.

CoilAdrH is 0
CoilAdrL is as follows:

0

Channel 1 Alarm (red) LED

1

Channel 2 Alarm (red) LED

2

Channel 3 Alarm (red) LED

3

Channel 4 Alarm (red) LED

4

Channel 1 Warning (Amber) LED

5

Channel 2 Warning (Amber) LED

6

Channel 3 Warning (Amber) LED

7

Channel 4 Warning (Amber) LED

8

Channel 1 OK (Green) LED

9

Channel 2 OK (Green) LED

10

Channel 3 OK (Green) LED

11

Channel 4 OK (Green) LED

12

Channel 1 Alarm Relay

13

Channel 2 Alarm Relay

14

Channel 3 Alarm Relay

15

Channel 4 Alarm Relay

16

Channel 1 Warning Relay

17

Channel 2 Warning Relay

18

Channel 3 Warning Relay

19

Channel 4 Warning Relay

20

Channel 1 OK Relay

21

Channel 2 OK Relay

22

Channel 3 OK Relay

23

Channel 4 OK Relay

24

System Alarm Relay