PrimeTest 350 Operating

Instructions

- 99 -

15.4 Protective Earth Conductor Current


Measurement Method

Differential

Test Voltage

Mains supply 230V +10% / -15%

Max Load Current

16A

Test Duration

2, 5, 10, 20, 30 seconds

Pass/Fail

Limit

User

Defined

Display

Range

0.00

19.99mA

Measuring Range

0.25 – 19.99mA

Resolution

0.01mA

Accuracy

±

10%

±2

digits

Attention: Avoid prolonged, repeated use at
full load (16A) and excessive test duration as

this may reduce the life of the unit.

15.5 Touch

Current


Test Voltage

Mains supply 230V +10% / -15%

Max Load Current

16A

Input Impedance

2K

Ÿ As per IEC 61010-1 A.1

Test Duration

2, 5, 10, 20, 30 seconds

Pass/Fail

Limit

User

Defined

Display

Range

0.00

3.50mA

Measuring Range

0.10 – 3.50mA

Resolution

0.01mA

Accuracy

±

10%

±2

digits

Attention: Avoid prolonged, repeated use at
full load (16A) and excessive test duration as

this may reduce the life of the unit.

15.6 Voltage/Frequency

Measurement


Incoming mains voltage and frequency are displayed
during the 2 powered tests only.

Voltage Display Range

0 – 300V

Voltage Measuring Range

20 – 300V

Voltage

Resolution

1V

Voltage

Accuracy

±

10%

Frequency Measuring Range

50Hz or 60Hz

Frequency

Accuracy

Indication

only