PrimeTest 350 Operating
Instructions
- 43 -
10.1.1 Asset ID
This can be set to automatically increment the ID
number at the end of each test, present a blank field or
Repeat Last ID.
10.1.2 Start Increment
Enter an Asset ID here, this will be used as the start
number for the auto increment feature.
10.1.3 On Test Failure
In the event of a test failure during an automatic
sequence, the PrimeTest 350 can be configured to
either End Test, which will terminate the test sequence
or Fail Menu, which will present a list options.
Fail menu gives the option to
restart the test, restart the
appliance (from the first test),
end the appliance (and record a
fail result) or abort the appliance.
Note: Restart Test is useful for coping with
situations such as forgetting to connect the
earth continuity probe.
End Appliance will record a Fail result
whereas Abort Appliance does not record a
result.
10.1.4 After Test
There is a choice of 4 actions to be carried out at the
end of a test sequence.