PrimeTest 350 Operating

Instructions

- 36 -

In the example Insulation Test above, the required
duration in seconds, pass/fail threshold in M

Ω and

number of times the test is to be repeated can be
entered using the keypad. Also the test voltage (250v
or 500v) can be selected using the left / right arrow
keys. When the required data has been entered, press
OK (F4) to accept.

The new test will now appear in the sequence with the
values entered for each parameter shown.

To add further tests, position the highlight bar at the
required point in the sequence and press Insert (F4).

Note: Tests are inserted into the sequence in the

position above the highlight bar.

9.1.5 User

Tests

The PrimeTest 350 has the unique ability to create
user-defined tests. When using the test sequence
editor, these appear in the list of available tests as
shown below.


The User test name and units of measurement are fully
configurable to allow the PrimeTest 350 to record both
electrical and non-electrical measurement data.

Note: The results of a User Test are stored within a

test record for post processing.

User Tests can be added to factory test
sequence or new user defined sequences.