Seaward 350 PrimeTest User Manual

Page 58

Advertising
background image

PrimeTest 350 Operating

Instructions

- 57 -

appliance to earth via the ‘body model’. The Tester
displays the result in milliamps (mA).
In the top left area of the screen, the tester also
displays the EUT load in kVA.


Warning: Always test the insulation
resistance before performing a Touch
Current test.

Warning: Always check that an appliance
with moving parts (e.g. an electric drill) is
safely mounted to avoid risk of damage to

equipment or personnel.


Attention: Avoid prolonged, repeated use at
full load (16A) and excessive test duration as
this may reduce the life of the unit.



Press Setup (F3) to set the Test Duration and
Pass/Fail limit. Use the arrow keys to highlight and set.
Press OK (F2) when complete. Esc (F2) will exit setup
mode without saving changes.

Fit the power lead supplied to the PrimeTest 350 and
plug the other end into a suitable mains outlet. Fit the
Appliance mains plug into the EUT outlet. Connect the
Earth Bond lead to the appliance metal part. Press the
green START key to begin the test.

Pre-load check - If the potential load current is too
high, a warning message appears, preventing the user
from continuing. This message will be displayed if a
Live to Neutral short exists or the EUT mains current
consumption is too high.

Internal Relay checks – Before applying power to the
EUT the PrimeTest 350 performs a series of internal
safety checks to verify that internal relays are working
properly.

Once the test has started it will run for the duration set.
To end the test sooner, press the red STOP key.