Ma/hart wiring, Basic ma output wiring, Hart/analog single loop wiring – Emerson MICRO MOTION 1700 User Manual

Page 100: 2 ma/hart wiring, 1 basic ma output wiring, 2 hart/analog single loop wiring

Advertising
background image

You cannot configure the combination of Channel B as discrete output and Channel C as
frequency output.

9.2

mA/HART wiring

9.2.1

Basic mA output wiring

Basic mA output wiring

Figure 9-1:

A

B

A.

820 Ω maximum loop resistance

B.

420 Ω maximum loop resistance

9.2.2

HART/analog single loop wiring

Note

For HART communications:

600 Ω maximum loop resistance

250 Ω minimum loop resistance

I/O wiring for Model 2700 transmitters with configurable input/outputs

96

Micro Motion

®

Model 1700 and 2700

Advertising