Bend radii of armored cable

Table 5-5:

Jacket material

Outside diameter

Minimum bend radii

Static (no load) condition

Under dynamic load

PVC

0.525 inches (14 mm)

4–1/4 inches (108 mm)

8–1/2 inches (216 mm)

Teflon FEP

0.340 inches (9 mm)

3–1/4 inches (83 mm)

6–3/8 inches (162 mm)

Cable illustrations

Cross-section view of jacketed cable

Figure 5-13:

A

C (4)

B (4)

D (5)

A.

Outer jacket

B.

Drain wire (4 total)

C.

Foil shield (4 total)

D.

Filler (5 total)

Mounting and sensor wiring for remote core processor with remote sensor installations