Components of a 4-wire remote mount transmitter (stainless steel

housing)

Figure 5-4:

A

B

D

C

A. End cap
B. Cap screws
C. Transmitter
D. Mounting bracket

2.

Attach the mounting bracket to an instrument pole.

5.2

Mount the remote core processor

This procedure is required only for remote core processor with remote transmitter
installations.

For mounting the remote core processor to a wall:

Use four 5/16-inch diameter (or M8) bolts and nuts that can withstand the process
environment. Micro Motion does not supply bolts or nuts (appropriate bolts and
nuts are available as an option).

Ensure that the surface is flat and rigid, does not vibrate, or move excessively.

For mounting the remote core processor to an instrument pole:

Use two 5/16-inch U-bolts for 2-inch pipe, and four matching nuts, that can
withstand the process environment. Micro Motion does not supply U-bolts or nuts
(appropriate bolts and nuts are available as an option).

Ensure the instrument pole extends at least 12 inches (305 mm) from a rigid base,
and is no more than 2 inches (50.8 mm) in diameter.

1.

If desired, reorient the core processor housing on the bracket.

Mounting and sensor wiring for remote core processor with remote sensor installations

Installation Manual